DLS Guide: Measuring Nanoparticle Size
Lab note: this article was written by AI. People can correct it, but it remains automated, so treat it as general information. Check methods, quantities and safety data against the original sources before you use them in a lab.
Dynamic Light Scattering (DLS), also known as Photon Correlation Spectroscopy, is a standard technique for measuring the hydrodynamic size of nanoparticles, proteins, and polymers in solution. In biotechnology and pharmaceutical development, DLS is indispensable for monitoring protein aggregation, verifying the success of nanodrug delivery systems, and ensuring the stability of viral vectors [1].
Unlike other sizing techniques that might require dehydration or complex calibration, DLS provides high-speed, non-destructive measurements of particles from about 0.5 nm to 2.5 µm in a single measurement [1]. This guide provides a deep dive into the mechanics of DLS, how to optimize your experimental setup, and how to avoid the common pitfalls that lead to inaccurate data.
Your reading tools
- SimplifyAny section in plain language
- SummarizeThe key points at a glance
- FAQQuick answers to common questions
- NewSpotted a mistake? We fix it in minutes!
Written with AI from the sources listed at the end.
Table of Contents
- How DLS Works: The Physics of Brownian Motion
- Step-by-Step Method Development
- Interpreting DLS Data: Intensity vs. Volume vs. Number
- Complementary Techniques
- Zeta Potential and Colloidal Stability
- Action Plan
- Sources
How DLS Works: The Physics of Brownian Motion
DLS does not measure particle size directly. Instead, it measures the Brownian motion of particles—the random, zig-zag movement caused by collisions with solvent molecules.
When a laser hits a sample, the light scatters in all directions. Because the particles are moving, the intensity of this scattered light fluctuates over time. Smaller particles move faster, causing rapid fluctuations, while larger particles move slowly, causing more gradual changes in light intensity [3].
The instrument records these fluctuations and uses a mathematical process called autocorrelation to determine the translational diffusion coefficient (Dt). This coefficient is then converted into a hydrodynamic radius (Rh) using the Stokes-Einstein equation:
Where:
-
kB: Boltzmann constant
-
T: Temperature
-
η: Solvent viscosity
Because DLS relies on the diffusion of a “hydrated” particle, the size it measures includes the water molecules moving with the particle. This is why DLS results are often slightly larger than those from Transmission Electron Microscopy (TEM), which measures the “dry” physical diameter; colloidal gold with a surfactant layer, for example, appears larger by DLS, which includes the layer, than by TEM [2].
No, DLS measures the hydrodynamic radius, which includes the particle and its surrounding layer of solvent molecules. This is why DLS results are typically larger than those obtained from dry measurement techniques like TEM.
Smaller particles move faster due to Brownian motion, causing rapid fluctuations in light intensity. These fluctuations are analyzed via the Stokes-Einstein equation to calculate the diffusion coefficient and subsequent particle size.
Step-by-Step Method Development
To get reliable data, you must treat DLS as a precise analytical method rather than a “black box” measurement. Malvern Panalytical’s method-development note covers instrument verification, sample preparation, measurement set-up and results analysis [5]; the steps below follow the same logic.
1. Sample Preparation and Filtration
The most common cause of poor DLS data is “dust” or large contaminants. For particles much smaller than the wavelength, scattered intensity rises with the sixth power of particle size [4] (r6): a particle ten times larger scatters about a million times more light, so a few dust particles can swamp the signal of the target nanoparticles. Filtration or centrifugation to remove dust is therefore critical [2].
- Best Practice: Filter your buffers to 0.02 µm or 0.1 µm and measure the pure buffer as a control. Centrifuge or filter the final sample only in a way that does not remove the analytes of interest, such as aggregates you may want to detect [1].
2. Concentration Optimization
There is a “Goldilocks” zone for sample concentration.
-
Too Low: There isn’t enough signal (count rate) to produce a smooth autocorrelation curve.
-
Too High: You encounter multiple scattering, where light bounces off multiple particles before hitting the detector, leading to an artificially small size reading.
-
Action: Perform a dilution series. Your measured size should remain constant across a range of concentrations. If the size changes as you dilute, you are likely seeing concentration-dependent interactions or multiple scattering; DLS results should be independent of sample concentration [5].
3. Setting Solvent Parameters
Since Rh is calculated using viscosity (η), entering the wrong solvent parameters will invalidate your results.
- Prescriptive Advice: If you are using a standard buffer (like PBS), ensure the software accounts for any additives (like glycerol or DMSO) that significantly alter viscosity. For more complex molecular interaction studies, consider checking out our ITC Guide: Measuring Molecular Binding Interactions to see how these environmental factors impact molecular behavior.
4. Verification with Standards
DLS cannot be calibrated in the traditional sense, but it must be verified. Use NIST-traceable polystyrene latex spheres to ensure the laser, detector, and software are performing according to specifications, prepared in 10 mM NaCl rather than deionized water, which would make them look larger [5].
For small particles, scattered intensity rises with the sixth power of particle size, so a few dust particles can overwhelm the signal from millions of nanoparticles. Using 0.02 µm or 0.1 µm filters is essential to remove these contaminants.
A concentration that is too high leads to multiple scattering, where light hits several particles before reaching the detector. To detect this, perform a dilution series; if the measured size changes as you dilute, the original sample was likely too concentrated.
Because the Stokes-Einstein equation uses viscosity to calculate size, any error in the solvent parameters will lead to a 1:1 error in your results. Always account for additives like glycerol or DMSO that change buffer viscosity.
Interpreting DLS Data: Intensity vs. Volume vs. Number
When you receive a DLS report, you will see three different distributions. Understanding the difference is critical for accurate reporting:
- Intensity Distribution: The raw data. It is inherently biased toward larger particles due to the r6 relationship. Even a small amount of aggregates can appear as a dominant peak; converted to a mass-weighted distribution, the same aggregates may be a small share [1].
- Volume Distribution: Uses Mie theory to estimate the relative volume of particles, which requires the refractive index and absorbance of the particle material [2]. This is often used when trying to determine the mass balance of a sample.
- Number Distribution: Estimates the actual number of particles. This is the most comparable to TEM data but is highly sensitive to errors in the initial intensity reading.
Which to use? In quality control, results are usually reported as the Z-Average (an intensity-weighted mean size) and the Polydispersity Index (PDI), which cumulants analysis estimates for a single-peak distribution [3]. For certified size standards, ISO 22412 requires a PDI below 0.1 in each repeat measurement [5]; much higher values point to a broad distribution that a single average describes poorly.
| Distribution Type | Primary Basis | Sensitivity Note |
|---|---|---|
| Intensity | Light Scattering (r6) | Highly biased toward large aggregates. |
| Volume | Mass/Physical Space | Useful for mass balance estimates. |
| Number | Particle Count | Comparable to TEM; sensitive to data noise. |
The intensity distribution is inherently biased toward larger particles because scattering intensity increases with the sixth power of the radius. Even a 1% presence of aggregates can dominate the intensity plot.
In pharmaceutical and quality control settings, the Z-Average and the Polydispersity Index (PDI) are the usual metrics. Size standards must show a PDI below 0.1 under ISO 22412; much higher values point to a broad distribution that a single average describes poorly.
Complementary Techniques
DLS is powerful but has limits, particularly with resolution. It struggles to distinguish two populations of particles unless their sizes differ by a factor of about 3 to 5 [1].
To overcome this, researchers often pair DLS with:
-
SEC-MALS: Size Exclusion Chromatography coupled with Multi-Angle Light Scattering. This physically separates the particles before measuring them, providing much higher resolution for polydisperse samples [3].
-
HPLC: Standard high-performance liquid chromatography is often the first step in sample purification. Learn more in our HPLC Guide for Small Molecule Analysis.
DLS struggle to resolve two different particle populations unless their sizes differ by a factor of about 3 to 5. SEC-MALS is preferred for complex or polydisperse samples because it physically separates the particles before measuring them.
Yes, DLS is an excellent tool for verifying if proteins remain monomeric or have begun to aggregate. However, pairing it with HPLC or SEC-MALS provides higher resolution and more definitive proof of sample purity.
Zeta Potential and Colloidal Stability
Every liquid dispersion, from paints to drug formulations, depends on a delicate balance of forces. Particles in a suspension are naturally drawn together by van der Waals forces. To prevent them from colliding and sticking together, they must possess a repulsive force—typically an electrostatic charge.
When a charged nanoparticle is placed in a liquid, it attracts a thin layer of ions with an opposite charge, known as the Stern layer. Surrounding this is a broader “diffuse layer” where ions are less tightly bound. The boundary where these ions begin to move independently of the particle is called the slipping plane. The electrical potential at this specific boundary is the Zeta Potential (ZP) [6].
Why the ±30 mV Threshold Matters
A zeta potential magnitude of about 30 mV (either +30 mV or -30 mV) is a common rule of thumb: magnitudes of 30–40 mV are classed as moderate stability, 40–60 mV as good and above 60 mV as excellent [7].
-
High Zeta Potential (> ±30 mV): Particles repel each other strongly, which helps keep the suspension dispersed.
-
Low Zeta Potential (below about ±30 mV): Magnitudes of 10–30 mV indicate incipient instability and 0–5 mV rapid coagulation or flocculation [7], as repulsion becomes too weak to overcome van der Waals attraction. Preventing aggregation matters for drug delivery systems such as liposomes and nanoparticles [6].
| Zeta Potential (mV) | Stability Classification |
|---|---|
| > ±60 mV | Excellent Stability |
| ±40 to ±60 mV | Good Stability |
| ±30 to ±40 mV | Moderate Stability |
| ±10 to ±30 mV | Incipient Instability |
| 0 to ±5 mV | Rapid Coagulation / Flocculation |
The slipping plane is the boundary positioned at the edge of the diffuse layer where ions move independently of the particle. The electrical potential measured at this specific interface is defined as the zeta potential.
Nanoparticles in suspension naturally attract each other via van der Waals forces; however, a high electrostatic surface charge creates a repulsive force that keeps particles apart, maintaining a uniform suspension.
The Stern layer consists of a thin layer of ions with an opposite charge that are tightly bound to the particle surface. It forms the first part of the electrical double layer that contributes to the overall zeta potential.
Action Plan
- Filter buffers: Use 0.02 µm or 0.1 µm filters for aqueous buffers [1].
- Check the Count Rate: On a Malvern Zetasizer Advance, the optimum count rate is 300–500 kcps where possible [5]; other instruments have their own ranges.
- Run in Triplicate: Perform at least three measurements to ensure the sample is stable and not aggregating over time.
- Validate: Measure a certified latex size standard at an interval you define to verify instrument performance [5].
DLS remains the fastest way to confirm your nanoparticles are the right size and that your proteins remain monomeric. By strictly controlling your sample environment and understanding the mathematical weighting of your data, you can turn a simple light scattering measurement into a robust analytical tool.
Malvern recommends 300–500 kcps where possible on its Zetasizer Advance; below 100 kcps at full laser power, a longer sub-run can improve data quality. Other instruments have their own ranges.
The lab defines the frequency: verify the instrument by measuring a standard of known size, such as NIST-traceable polystyrene latex spheres, at regular intervals. This ensures the laser, detector, and software are performing according to the manufacturer’s specifications.
Sources
- [1] Wyatt Technology: DLS Overview
- [2] Dynamic light scattering – Wikipedia
- [3] Wyatt Technology: Measuring Size with Light Scattering
- [4] Rayleigh scattering – Wikipedia
- [5] Malvern Panalytical: DLS Method Development
- [6] Wyatt Technology: Zeta Potential Overview – Fundamentals and Applications
- [7] Zeta potential – Wikipedia