A Practical Guide to Sample Preparation for Infrared Spectroscopy (ATR, KBr, Nujol)
Lab note: this article was written by AI. People can correct it, but it remains automated, so treat it as general information. Check methods, quantities and safety data against the original sources before you use them in a lab.
Infrared (IR) spectroscopy remains one of the most vital tools in the analytical chemist’s arsenal for identifying functional groups and verifying molecular structures. However, the quality of an IR spectrum is almost entirely dependent on how the sample is presented to the spectrometer. A poorly prepared sample results in “bottomed out” peaks, scattering noise, or atmospheric interference that can render data useless.
This guide provides a technical walkthrough of the three primary sampling techniques used in Modern Fourier Transform Infrared (FTIR) labs: Attenuated Total Reflectance (ATR), Potassium Bromide (KBr) pellets, and Nujol mulls. To better understand how these results are processed, you may also want to refer to our practical guide to FTIR spectroscopy and functional group interpretation.
Your reading tools
- SimplifyAny section in plain language
- SummarizeThe key points at a glance
- FAQQuick answers to common questions
- NewSpotted a mistake? We fix it in minutes!
Written with AI from the sources listed at the end.
Table of Contents
- 1. Attenuated Total Reflectance (ATR)
- 2. The KBr Pellet Method
- 3. Nujol Mull (Liquid Film)
- Troubleshooting Common Preparation Issues
- Summary of Key Takeaways
- Sources
1. Attenuated Total Reflectance (ATR)
ATR has largely replaced traditional transmission methods in modern laboratories due to its speed and minimal cleanup. Instead of the IR beam passing through the sample, it reflects off the internal surface of a high-refractive-index crystal (typically diamond, Zinc Selenide, or Germanium) in contact with the sample [1].
When to Choose ATR
-
Best for: Rapid screening, liquids, pastes, polymers, and hard solids.
-
Avoid for: Samples with very low concentrations, materials that might scratch a soft crystal such as ZnSe, and layered or uneven materials: ATR mostly sees the region near the surface, which may not match the bulk [3].
Step-by-Step Procedure
- Clean the Crystal: Use a lint-free wipe with isopropanol or acetone. Perform a “Background” scan of the clean, open crystal to subtract atmospheric CO₂ and water vapor.
-
Apply the Sample:
-
Liquids: Place one drop directly over the crystal area.
-
Solids: Place a small amount of powder or a piece of the material on the center of the crystal.
- Apply Pressure: For solids, use the built-in pressure clamp. According to JASCO, the IR beam penetrates only about 1.5 micrometers into the sample with a diamond crystal, and 0.8 micrometers with germanium [1]. Without firm contact, the signal will be weak or non-existent.
- Acquire and Correct: The penetration depth depends on the refractive indices of the sample and the crystal, and so on the wavelength: an ATR spectrum has slightly different intensity ratios from a transmission spectrum (bands at lower wavenumbers come out stronger), and “ATR Correction” in your software corrects them for comparison with transmission spectra [1].
-
ATR is significantly faster and requires minimal cleanup because it eliminates the need for complex sample preparation like making KBr pellets or Nujol mulls. It allows for the direct analysis of liquids, pastes, and solids by simply placing the sample in contact with a high-refractive-index crystal.
The IR beam penetrates only about 1.5 micrometers into the sample with a diamond crystal (0.8 micrometers with germanium), so intimate contact between the sample and the crystal is required. Without applying sufficient pressure using a clamp, the IR signal will be too weak to produce a usable spectrum.
Avoid ATR for samples with very low concentrations that require the higher sensitivity of transmission methods, or when using materials that could scratch sensitive crystals like Zinc Selenide (ZnSe).
2. The KBr Pellet Method
The KBr pellet technique is the traditional “gold standard” for transmission IR. Potassium Bromide is used because it is transparent in the mid-IR range (4,000–400 cm⁻¹) and forms pellets under high pressure [1], holding the sample in the beam as a solid, transparent matrix [2].
When to Choose KBr Pellets
-
Best for: Powdered solids, even when you have only a small amount, measured in transmission with little light scattering [3].
-
Avoid for: Samples with bands in the same regions as water (KBr is itself hygroscopic and picks up moisture), and chlorides or hydrochlorides, which exchange halogen with the bromide: press those in KCl instead [1].
Step-by-Step Procedure
- Preparation (Dilute the Sample): Aim for 0.2 to 1 percent sample in dry KBr (about 0.4 to 2 mg in 200 mg). The pellet is much thicker than a liquid film, so a low concentration is needed; too much sample gives cloudy pellets and bands that absorb completely [4].
- Grinding: Use an agate mortar and pestle to grind the mixture until it has the consistency of fine flour. If the particles are too large, light is scattered out of the beam (Christiansen scattering) and the bands come out distorted; do not over-grind the KBr itself, because the finer it is, the more water it picks up from the air [4].
- Pressing: Transfer the powder to a pellet die. Compress it in a hydraulic press: under a load of several tons the KBr flows and fuses into a glass-like, transparent disc with the sample trapped inside [2]. Use the load and time your die and press are made for.
- Analysis: The resulting pellet should be nearly transparent. Place it in a universal holder for transmission analysis.
Tip: A cloudy pellet can come from a mixture not ground enough, a sample that was not dry, too much sample, a pellet that is too thick or a sample with a low melting point. Keep the KBr dry (NIU’s lab keeps it in an oven at 100°C) and work fast [4].
KBr is used because it is transparent to infrared radiation in the mid-IR range (4000–400 cm⁻¹), allowing the beam to pass through the sample without the carrier medium interfering with the spectra.
A KBr pellet is much thicker than a liquid film, so the sample must be dilute: NIU’s lab gives 0.2 to 1 percent sample in KBr. Too much sample makes the pellet cloudy and the bands absorb completely, which gives noisy spectra.
Cloudy disks have several causes: a mixture not ground enough, a sample that was not dry, too much sample, a pellet that is too thick, or a sample with a low melting point. Keep the KBr dry (NIU’s lab stores it in an oven at 100°C) and work quickly, because finely powdered KBr absorbs moisture from the air.
3. Nujol Mull (Liquid Film)
A “mull” is a suspension of a solid in a heavy mineral oil (Nujol). This is a legacy technique still used when ATR is unavailable or when a sample is sensitive to the high pressures of KBr pressing.
When to Choose Nujol
-
Best for: Quick transmission checks of solids when a hydraulic press is unavailable, and solids that do not suit the KBr method, such as samples that absorb moisture during preparation (sugars, for example) or that do not dissolve in solvents (inorganic substances) [1].
-
Avoid for: Samples whose C-H bands matter: Nujol is a liquid paraffin, which absorbs at 3,000–2,800 cm⁻¹ and 1,500–1,300 cm⁻¹ and hides the sample’s own bands there [1].
Step-by-Step Procedure
- Grind the Solid: Grind 5 to 10 mg of sample in a mortar until the particles are about 1 to 2 microns, smaller than the wavelengths they must let through.
- Add the Oil: Place the ground sample on the face of a salt plate (NaCl or KBr) and add a small drop of mineral oil.
- Sandwich: Place the second plate on top and spread the mixture with a gentle circular, back-and-forth rubbing motion; a good mull looks slightly translucent, with no bubbles [4].
- Clean Up: Salt plates dissolve or fog up with water. NIU’s lab wipes them with a tissue, washes them several times with methylene chloride and then ethanol, polishes them, and handles them with gloves to prevent fogging [4].
Nujol is a liquid paraffin, so it absorbs strongly in the C-H regions (3,000–2,800 cm⁻¹ and 1,500–1,300 cm⁻¹) and hides your sample’s own bands there.
Salt plates (NaCl or KBr) dissolve or fog up with water, so never clean them with water or aqueous solutions. NIU’s lab wipes them with a tissue, washes them several times with methylene chloride and then ethanol, polishes them and handles them with gloves to prevent fogging.
Troubleshooting Common Preparation Issues
| Symptom | Probable Cause | Solution |
|---|---|---|
| Flat-topped peaks | Sample is too thick/concentrated | Use less sample (KBr) or a thinner film (Mull) |
| Noisy baseline | Poor contact (ATR) or scattering (KBr) | Increase clamp pressure or grind sample finer |
| Huge broad peak at 3400 cm⁻¹ | Water contamination | Dry your KBr; avoid breathing on the sample |
| Shifted Baselines | Refractive index mismatch | Ensure sample is finely ground to reduce light scattering |
Distorted peaks, often called the Christiansen Effect, occur when the sample particles are not ground finely enough. Ensuring the sample-KBr mixture is ground to the consistency of fine flour in an agate mortar can mitigate this issue.
A background scan accounts for current environmental factors like atmospheric CO₂ and water vapor. Subtracting this background ensures that the final spectrum only represents the chemical structure of your sample rather than the air inside the spectrometer.
Summary of Key Takeaways
| Method | Sample Physics | Primary Advantage |
|---|---|---|
| ATR | Surface Reflection | No preparation; non-destructive |
| KBr Pellet | Transmission | Little light scattering; works with small amounts |
| Nujol Mull | Transmission | For solids that do not suit KBr |
Decision Matrix: Which Method to Use?
-
Use ATR for most routine lab work: JASCO calls it the primary sampling method for FTIR, because liquids and solids need no preparation [1]. It is non-destructive, requires no salt plates, and is the fastest method for liquids and solids.
-
Use KBr Pellets when you need high-quality transmission spectra for publication or when looking for very weak signals that ATR might miss.
-
Use Nujol only if you do not need the C-H regions and want a quick transmission alternative for a solid powder.
Action Plan for Beginners
- Identify your Regions of Interest: If you need to see the 3000 cm⁻¹ region, skip Nujol.
- Check Sample Solubility: If your solid is soluble in a volatile solvent, you can “cast” a thin film directly onto a salt plate instead of making a pellet.
- Perform a Background Check: Always run a background scan before every sample to account for changes in lab humidity and CO₂.
- Clean Promptly: Salt plates and ATR crystals can be etched by acidic or aqueous samples. Clean them immediately after your scan is complete.
IR spectroscopy is as much an art as it is a science. By mastering these three preparation techniques, you ensure that your data is a true reflection of the molecular structure rather than an artifact of poor lab technique.
ATR is the usual choice for routine work: liquids and solids need no preparation, it is non-destructive, requires no special salt plates, and is the fastest method for both liquids and solids.
Use ATR for speed and convenience with most solids and liquids. Choose KBr pellets when you need high-quality transmission data for publication, and use Nujol only as a quick transmission alternative if you don’t need to see the C-H regions.